1. Scope of Application
The stability study of passive implanted medical devices and the preparation of relevant registration application materials. The stability study generally includes shelf life, service stability, and transport stability.
2.1 Shelf Life
The verification test of the shelf life of medical devices can usually be divided into two categories: accelerated stability tests and real-time stability tests.
Accelerated Stability Test
- AAT: Accelerated Aging Time.
- RT: Real-Time Aging Time.
- Q10: Aging coefficient increases or decreases by 10°C; Q10 is generally set to 2, but can also be conservatively adjusted.
- TAA: Accelerated Aging Temperature. Generally, the accelerated aging temperature should not exceed 60°C.
- TRT: Preset current storage temperature. The relative humidity is generally recommended to be 60% ±10%, and it is recommended to verify the performance of the product under the most adverse conditions.
2.2 Stability of Use
To maintain its safety and effectiveness through the life cycle of normal use, maintenance (if applicable) is necessary.
2.3 Transportation Stability
Transportation stability is usually assessed through simulated transportation tests. By simulating changes in the environment (temperature, humidity, air pressure, etc.), drop, collision, vibration, and acceleration during transportation, it is ensured that the product will not change in performance and function, and that the packaging system has the ability to protect the product.
New link: https://www.cmde.org.cn/index.html
Type Approval impacted: MEDICAL
Date of published: 04/27/22